Byli jste přesměrováni na vaši místní verzi požadované stránky

Find more information about maintenance intervals depending on the interval periods (2 months, quarterly, semi-annually, or annually), sample used, or topic (IC system, ultrafiltration, dialysis, PCR, columns, miscellaneous).

Interval – 2 months

Description Order number
Peristaltic pump tubing (orange/yellow) with three stoppers for MSM regeneration solution 6.1826.420
Peristaltic pump tubing (white/white) with three stoppers to pump the sample from the Sample Processor to the injection loop 6.1826.360
Peristaltic pump tubing (yellow/yellow) with three stoppers used on the sample side of the Ultrafiltration cell 6.1826.390
Peristaltic pump tubing (orange/yellow) with three stoppers used on the filtrate side of the Utlrafiltration cell 6.1826.320
Peristaltic pump tubing (black/black) with three stoppers used to pump the sample solution for dialysis 6.1826.340
Peristaltic pump tubing (orange/yellow) with three stoppers used to pump the acceptor solution for dialysis 6.1826.320
Peristaltic pump tubing (orange/green) with three stoppers used to pump the PCR reagent for the bromate determination with the triiodide method 6.1826.310

Interval – Quarterly

Description Order number
Aspiration filter for the eluent
(set of 5)
6.2821.090
Spare filter for inline filter (6.2821.120) between purge valve and pulsation absorber
(set of 10)
6.2821.130
Spare filter for inline filter (6.2744.180) of the MSM 6.2821.130
Metrosep C PCC 1/4.0 6.1010.300
Metrosep C PCC 1 HC/4.0 6.1010.310
Metrosep C PCC 1 VHC/4.0 6.1010.320
Metrosep I Trap 1 - 100/4.0 6.1014.200
Metrosep A PCC 2/4.0 6.1006.330
Metrosep A PCC 2 HC/4.0 6.1006.340
Metrosep A PCC 2 VHC/4.0 6.1006.350

Interval – Semi-annually

Description Order number
940/930/925 IC Pump
Flow check
 -  
Metrosep C Trap 1 - 100/4.0 6.1015.000
Metrosep A Trap 1 - 100/4.0 6.1014.000

Interval – Annually

Description Order number
940/930/925 IC Pump
Inlet valve PEEK
6.2824.170
940/930/925 IC Pump
Outlet valve PEEK
6.2824.160
940/930/925 IC Pump
Piston seal (orange)
6.2741.020
940/930/925 IC Pump
Zirconium oxide piston
6.2824.070

Complete glass cylinder unit, 50 mL, for Inline Eluent Preparation

Complete glass cylinder unit, 2 mL, for Dosino Regeneration

6.1574.250

6.1574.120

CO2 adsorption cartridge CW 6.2837.100

Interval – Depending on sample

Description Recommended maintenance interval Order number
Metrosep A Supp 4/5 Guard/4.0 (example guard)* The use of 3-4 guards per analytic column is recommended. 6.1006.500
Metrosep C 4 Guard /4.0 (example guard)* The use of 3-4 guards per analytic column is recommended. 6.1050.500
Metrosep A Supp 5 - 150/4.0 (example column)* 2500 injections, highly sample depending 6.1006.520
Metrosep C 4 - 150/4.0 (example column)* 2500 injections, highly sample depending 6.1050.420
Ultrafiltration membrane
(set of 50)
>100 injections, depending on sample matrix 6.2714.020
Dialysis membrane cellulose (standard)
(set of 50)
40-400 injections, depending on sample matrix 6.2714.010
Dialysis membrane polyamide (alternative)
(set of 50)
40-400 injections, depending on sample matrix 6.2714.030
Vials PP, 11 mL (2000 pieces) For single use 6.2743.050
Vials PP, 11 mL (200 pieces) For single use 6.2743.057
Vials PP, 2.5 mL (2000 pieces) For single use 6.2743.040
Vials PP, 2.5 mL (200 pieces) For single use 6.2743.047
Stopper with perforation (2000 pieces) For single use 6.2743.070
Stopper with perforation (200 pieces) For single use 6.2743.077

*Caution: Number of injections can vary! It is highly depending on the sample and on the care of use!

Topic – IC system

Description Recommended maintenance interval Order number
Peristaltic pump tubing (orange/yellow) with three stoppers for MSM regeneration solution 2 months 6.1826.420
Peristaltic pump tubing (white/white) with three stoppers to pump the sample from the Sample Processor to the injection loop 2 months 6.1826.360
Aspiration filter for the eluent
(set of 5)
quarterly 6.2821.090
Spare filter for inline filter (6.2821.120) between purge valve and pulsation absorber
(set of 10)
quarterly 6.2821.130
Spare filter for inline filter (6.2744.180) of the MSM quarterly 6.2821.130
940/930/925 IC Pump
Flow check
semi-annually -
940/930/925 IC Pump
Inlet valve PEEK
annually 6.2824.170
940/930/925 IC Pump
Outlet valve PEEK
annually 6.2824.160
940/930/925 IC Pump
Piston seal (orange)
annually 6.2741.020
940/930/925 IC Pump
Zirconium oxide piston
annually 6.2824.070
CO2 adsorption cartridge CW annually 6.2837.100

Topic – Ultrafiltration

Description Recommended maintenance interval Order number
Peristaltic pump tubing (yellow/yellow) with three stoppers used on the sample side of the Ultrafiltration cell 2 months 6.1826.390
Peristaltic pump tubing (orange/yellow) with three stoppers used on the filtrate side of the Utlrafiltration cell 2 months 6.1826.320
Ultrafiltration membrane
(set of 50)
>100 injections, depending on sample matrix 6.2714.020

Topic – Dialysis

Description Recommended maintenance interval Order number
Peristaltic pump tubing (black/black) with three stoppers used to pump the sample solution for dialysis 2 months 6.1826.340
Peristaltic pump tubing (orange/yellow) with three stoppers used to pump the acceptor solution for dialysis 2 months 6.1826.320
Dialysis membrane cellulose (standard)
(set of 50)
40-400 injections, depending on sample matrix 6.2714.010
Dialysis membrane polyamide (alternative)
(set of 50)
40-400 injections, depending on sample matrix 6.2714.030

Topic – PCR

Description Recommended maintenance interval Order number
Peristaltic pump tubing (orange/green) with three stoppers used to pump the PCR reagent for the bromate determination with the triiodide method 2 months 6.1826.310

Topic – Columns

Description Recommended maintenance interval Order number
Metrosep C PCC 1/4.0 quarterly 6.1010.300
Metrosep C PCC 1 HC/4.0 quarterly 6.1010.310
Metrosep C PCC 1 VHC/4.0 quarterly 6.1010.320
Metrosep I Trap 1 - 100/4.0 quarterly 6.1014.200
Metrosep A PCC 2/4.0 quarterly 6.1006.330
Metrosep A PCC 2 HC/4.0 quarterly 6.1006.340
Metrosep A PCC 2 VHC/4.0 quarterly 6.1006.350
Metrosep C Trap 1 - 100/4.0 semi-annually 6.1015.000
Metrosep A Trap 1 -100/4.0 semi-annually 6.1014.000
Metrosep A Supp 4/5 Guard/4.0 (example guard)* The use of 3-4 guards per analytic column is recommended. 6.1006.500
Metrosep C 4 Guard/4.0 (example guard)* The use of 3-4 guards per analytic column is recommended. 6.1050.500
Metrosep A Supp 5 -150/4.0 (example column)* 2500 injections, highly sample depending 6.1006.520
Metrosep C 4 -150/4.0 (example column)* 2500 injections, highly sample depending 6.1050.420

*Caution: Number of injections can vary! It is highly depending on the sample and on the care of use

Topic – Miscellaneous

Description Recommended maintenance interval Order number

Complete glass cylinder unit, 50 mL, for Inline Eluent Preparation

Complete glass cylinder unit, 2 mL, for Dosino Regeneration

annually

6.1574.250

6.1574.120

Vials PP, 11 mL (2000 pieces) For single-use 6.2743.050
Vials PP, 11 mL (200 pieces) For single-use 6.2743.057
Vials PP, 2.5 mL (2000 pieces) For single-use 6.2743.040
Vials PP, 2.5 mL (200 pieces) For single-use 6.2743.047
Stopper with perforation (2000 pieces) For single-use 6.2743.070
Stopper with perforation (200 pieces) For single-use 6.2743.077