Bismuth is considered as a metal with a very low toxicity. However, in high concentrations toxic effects have been described. There is no guideline value for bismuth in the World Health Organization’s «Guidelines for Drinking-water Quality» because typical levels usually found in drinking water are of no concern.
Anodic stripping voltammetry is a viable, less sophisticated alternative to atomic absorption spectroscopy (AAS) for the determination of bismuth in drinking water. While AAS (and competing methods) can only be performed in a laboratory, anodic stripping voltammetry can be used in the laboratory or alternatively in the field with the 946 Portable VA Analyzer. The determination is carried out on the scTRACE Gold electrode.
Bottled mineral water
The scTRACE Gold is electrochemically activated prior to the first determination. In the next step, the water sample and the supporting electrolyte are pipetted into the measuring vessel. The determination is carried out with the 884 Professional VA or with the 946 Portable VA Analyzer using the parameters specified in Table 1. The concentration is determined by two additions of a standard addition solution.
Parameter | Setting |
---|---|
Mode | DP – Differential Pulse |
Deposition potential | -0.1 V |
Deposition time | 30 s |
Start potential | -0.2 V |
End potential | 0.3 V |
Peak potential Bi | 0.08 V |
- scTRACE Gold
The limit of detection of the method is about 1 μg/L with the 946 Portable VA Analyzer, and approximately 0.03 μg/L with the 884 Professional VA.
Sample | Bi (μg/L) |
---|---|
Tap water | 20.4 |
Internal references: AW VA CH4-0568-042018, AW VA CH4-0577-022019