Edible oils are essential for our diet. Various parameters are used to assess oil quality including the determination of iodine value, free fatty acids (FFA), refractive index, and fatty acid composition. Fatty acid composition analysis provides a detailed view regarding the concentration of different fatty acids present in the oil. The content of essential linoleic acid (C18:2) and alpha-linolenic acid (C18:3) are especially interesting for edible oil producers.
Traditional analysis techniques like titration or gas chromatography can be time consuming and often require the use of hazardous solvents which can pose health risks and increase analysis costs. In contrast to these standard methods, each of the mentioned edible oil quality parameters can be analyzed simultaneously, without sample preparation, and in a few seconds via near-infrared spectroscopy (NIRS) with the OMNIS NIRS Analyzer.
More than 1000 samples of several types of edible oils (including sunflower-, rapeseed-, sesame-, and soybean oil) were measured on the OMNIS NIR Analyzer Liquid in transmission mode (1000–2250 nm) using 8 mm disposable vials. The temperature control of the NIRS analyzer was set to 40 °C to ensure consistent measurement performance. The OMNIS software was used for all data acquisition and prediction model development.
Equipment | Article number |
---|---|
OMNIS NIR Analyzer Liquid | 2.1070.0010 |
Holder OMNIS NIR, vial, 8 mm | 6.07401.070 |
Disposable vial, 8 mm, transmission | 6.7402.240 |
OMNIS Stand-Alone license | 6.06003.010 |
Quant Development software license | 6.06008.002 |
The obtained NIR spectra (Figure 2) were used to create prediction models to quantify all eight oil quality parameters: iodine value (IV), FFA, refractive index, palmitic acid (C16:0), stearic acid (C18:0), oleic acid (C18:1), linoleic acid (C18:2), and alpha-linolenic acid (C18:3). The quality of the prediction models was evaluated using correlation diagrams which display a high correlation (R2 > 0.94) between the NIR prediction and the standard reference methods for all parameters. Out of the total, 25% of the samples were selected as the validation set and the other 75% as a calibration set. The respective figures of merit (FOM) display the expected precision and confirm the feasibility during routine analysis (Figures 3–10).
Parameter | SEC (mg/100g) | SECV (mg/100g) | SEP (mg/100g) | R²CV |
---|---|---|---|---|
IV | 0.47 | 0.48 | 0.50 | 0.999 |
Parameter | SEC (%) | SECV (%) | SEP (%) | R²CV |
---|---|---|---|---|
FFA | 0.12 | 0.12 | 0.13 | 0.946 |
Parameter | SEC (%) | SECV (%) | SEP (%) | R²CV |
---|---|---|---|---|
RI | 0.00011 | 0.00012 | 0.00012 | 0.998 |
Parameter | SEC (%) | SECV (%) | SEP (%) | R²CV |
---|---|---|---|---|
C16:0 | 0.26 | 0.27 | 0.31 | 0.958 |
Parameter | SEC (%) | SECV (%) | SEP (%) | R²CV |
---|---|---|---|---|
C18:0 | 0.26 | 0.27 | 0.25 | 0.936 |
Parameter | SEC (%) | SECV (%) | SEP (%) | R²CV |
---|---|---|---|---|
C18:1 | 0.64 | 0.67 | 0.71 | 0.999 |
Parameter | SEC (%) | SECV (%) | SEP (%) | R²CV |
---|---|---|---|---|
C18:2 | 0.63 | 0.77 | 0.84 | 0.999 |
Parameter | SEC (%) | SECV (%) | SEP (%) | R²CV |
---|---|---|---|---|
C18:3 | 0.32 | 0.36 | 0.36 | 0.989 |
This Application Note displays the benefits of using the OMNIS NIR Analyzer Liquid for routine analysis of several QC parameters in the laboratories of edible oil manufacturers. All quality parameters can be measured simultaneously in only a few seconds. Compared to other conventional methods, measurements performed with NIR spectroscopy do not need any sample preparation or solvents. This ultimately leads to a reduction in workload (Table 2) and related costs, as well as keeping lab personnel safer.
Table 2. Time to result overview for the measurement of iodine value, FFA content, refractive index, and fatty acid composition in edible oils by standard analytical methods.
Parameter | Method | Time to result |
---|---|---|
Iodine value | Gas chromatography | ∼30 sample preparation (Methyl esterification + sample preparation) + ∼20 GC |
FFA content | Titration | ∼10 minutes per sample |
Refractive index | Refractometer | ∼2 minutes per sample |
Fatty acid composition | Gas chromatography | ∼30 minutes per sample |
Internal reference: AW NIR CH-0074-042023