You have been redirected to your local version of the requested page

The determination of sodium with the sodium ISE represents a selective, rapid, accurate, and favorably-priced method which is described in this Bulletin. Examples are used to show how determinations can be carried out with the 692 pH/Ion Meter using either direct measurement or the standard addition technique. The sodium concentration has been determined in standard solutions, water samples (tap water, mineral water, wastewater), foodstuffs (spinach, baby food), and urine. The construction, working principles, and areas of application of the two Metrohm ion-selective sodium electrodes – the 6.0501.100 Glass membrane ISE and the 6.0508.100 Polymer membrane ISE – are explained in detail.

Contact

Metrohm SA

1 Bridle Close
2052 Woodmead

Contact