Light naphtha isomerization is used by refineries to produce high-octane isomerate products which meet current gasoline specifications. Isomerization increases the octane value of light naphtha by increasing the degree of branching of paraffin molecules. The research octane number (RON) of the target product is dependent on various plant production parameters (e.g., temperature or hydrogen to hydrocarbon ratio). To optimize the plant process, a reliable and quick analytical method is key.
The standard method to determine RON in isomerate is with expensive and maintenance-intensive engines. In contrast to this, the research octane number can also be analyzed by near-infrared spectroscopy (NIRS). NIRS provides accurate results within one minute without the need for any sample preparation or chemicals.
63 different isomerate samples with varying RON values were measured with the Metrohm DS2500 Liquid Analyzer (Figure 1) in transmission mode over the full wavelength range of 400–2500 nm. The built-in temperature controller ensured measurement stability with a constant sample temperature of 35 °C. For convenience, disposable vials with a pathlength of 8 mm were used which made a cleaning procedure obsolete. The Vision Air Complete software package from Metrohm was used for data acquisition and prediction model development.
Table 1. Hardware and software equipment overview.
Equipment | Article number |
---|---|
DS2500 Liquid Analyzer | 2.929.0010 |
DS2500 Holder 8 mm vials | 6.7492.020 |
Vision Air 2.0 Complete | 6.6072.208 |
The obtained Vis-NIR spectra (Figure 2) were used to create a prediction model for the quantification of research octane number in isomerate. The quality of the prediction models was evaluated using correlation diagrams based on the cross-validation algorithm. A value of R2 > 0.98 displays a high correlation between the Vis-NIR prediction and the reference ASTM method. The respective figures of merit (FOM) display the expected precision during routine analysis (Figure 3).
Figures of Merit | Value |
---|---|
R2 | 0.986 |
Standard Error of Calibration | 0.73 |
Standard Error of Cross-Validation | 0.76 |
This Application Note demonstrates the feasibility of NIR spectroscopy for the analysis of RON in isomerate samples. In comparison to the conventional method, the time to result (Table 2) is a major advantage of Vis-NIR spectroscopy. With NIRS, a single measurement is performed within one minute, while the CFR engine test is much longer.
Table 2. Time to result for the determination of RON value with the standard reference method ASTM D2699.
Parameter | Method | Time to result |
---|---|---|
RON | CFR engine test | ∼30 minutes per sample |