Zostałaś(eś) przekierowany do lokalnej wersji strony

The most effective hand sanitizers contain between 62–95% alcohol. Alcohols are effective against most vegetative forms of bacteria, fungi, and enveloped viruses, but are ineffective against bacterial spores. The addition of hydrogen peroxide (3%) to the product may solve this, but due to its corrosive nature it must be handled with caution during production. Additionally, water and small amounts of emollient (e.g. glycerol) are added to protect the skin. Depending on the exact percentage of these constituents, hand sanitizer is either found in a liquid or gel form. Determination of the concentrations of these reagents is typically performed with gas chromatography (for glycerol and ethanol), Karl Fischer titration (for water), and redox titration (for H2O2). The disadvantage is that two different methods are needed which are time-consuming and require chemical reagents. Near-infrared spectroscopy (NIRS) on the other hand allows for the rapid and reliable simultaneous quantification of ethanol, glycerol, hydrogen peroxide, and water content in hand sanitizer formulations. 

DS2500 Liquid Analyzer and a sample filled in a  disposable vial.
Figure 1. DS2500 Liquid Analyzer and a sample filled in a disposable vial.

A total of 98 samples of hand sanitizer gel with different concentrations of glycerol (0.5–3 w/w %), ethanol (70–85 w/w %), and water (20–22 w/w %) were collected to create a prediction model for quantification. A total of 91 samples of liquid hand sanitizer used for sanitizing wipes were measured with different concentrations of ethanol (70–95 w/w %), water (2–40 w/w %), and hydrogen peroxide (0–4 w/w %). All samples were measured with a DS2500 Liquid Analyzer in transmission mode (400–2500 nm). Reproducible spectrum acquisition was achieved using the built-in temperature control at 40 °C: For convenience, disposable vials with a pathlength of 8 mm were used, which made cleaning of the sample vessels unnecessary. The Metrohm software package Vision Air Complete was used for all data acquisition and prediction model development.

Table 1. Hardware and software equipment overview
Equipment Metrohm number
DS2500 Liquid Analyzer 2.929.0010
DS2500 Holder 8 mm vials  6.7492.020
Disposable vials, 8 mm  6.7402.000
Vision Air 2.0 Complete  6.6072.208

All measured Vis-NIR spectra (Figure 2) were used to create a prediction model for quantification of the key quality parameters of gel and liquid sanitizer formulations. The quality of the prediction model was evaluated using correlation diagrams, which display a very high correlation between the Vis-NIR prediction and the reference values. The respective figures of merit (FOM) display the expected precision of a prediction during routine analysis. 

Figure 2. Vis-NIR spectra hand sanitizer gel samples analyzed on a DS2500 Liquid Analyzer.

Ethanol content in gel

Figure 3. Correlation diagram for the prediction of ethanol content in hand sanitizer gel using a DS2500 Liquid Analyzer. The lab value was evaluated using gas chromatography.
Table 2. Figures of merit for the prediction of ethanol content in hand sanitizer gel using a DS2500 Liquid Analyzer.
Figures of Merit Value
R2 0.9832
Standard Error of Calibration 0.33 w/w%
Standard Error of Cross-Validation 0.37 w/w%

Glycerol content in gel

Figure 4. Correlation diagram for the prediction of glycerol content in hand sanitizer gel using a DS2500 Liquid Analyzer. The lab value was evaluated by gas chromatography.
Table 3. Figures of merit for the prediction of glycerol content in hand sanitizer gel using a DS2500 Liquid Analyzer.
Figures of Merit Value
R2 0.9632
Standard Error of Calibration 0.08 w/w%
Standard Error of Cross-Validation 0.11 w/w%

Water content in gel

Figure 5. Correlation diagram for the prediction of water content in hand sanitizer gel using a DS2500 Liquid Analyzer. The lab value was evaluated by Karl Fischer titration.
Table 4. Figures of merit for the prediction of water content in hand sanitizer gel using a DS2500 Liquid Analyzer.
Figures of Merit Value
R2 0.941
Standard Error of Calibration 0.07 w/w%
Standard Error of Cross-Validation 0.09 w/w%

Ethanol content in liquid

Figure 6. Correlation diagram for the prediction of ethanol content in hand sanitizer wipes using a DS2500 Liquid Analyzer. The lab value was evaluated by gas chromatography.
Table 5. Figures of merit for the prediction of ethanol content in hand sanitizer wipes using a DS2500 Liquid Analyzer.
Figures of Merit Value
R2 0.9964
Standard Error of Calibration 0.36 w/w%
Standard Error of Cross-Validation 0.36 w/w%

Water content in liquid

Figure 7. Correlation diagram for the prediction of water content in hand sanitizer wipes using a DS2500 Liquid Analyzer. The lab value was evaluated by Karl Fischer titration.
Table 6. Figures of merit for the prediction of water content in hand sanitizer wipes using a DS2500 Liquid Analyzer.
Figures of Merit Value
R2 0.9999
Standard Error of Calibration 0.12 w/w%
Standard Error of Cross-Validation 0.18 w/w%

Hydrogen peroxide content in liquid

Figure 8. Correlation diagram for the prediction of hydrogen peroxide content in hand sanitizer wipes using a DS2500 Liquid Analyzer. The lab value was evaluated by permanganate titration.
Table 7. Figures of merit for the prediction of hydrogen peroxide content in hand sanitizer wipes using a DS2500 Liquid Analyzer.
Figures of Merit Value
R2 0.9986
Standard Error of Calibration 0.05 w/w%
Standard Error of Cross-Validation 0.06 w/w%

This application note demonstrates the feasibility to determine multiple key parameters of the quality control of liquid and gel-type hand sanitizer products with NIR spectroscopy. Vis-NIR spectroscopy enables a fast alternative to primary methods with high accuracy, and therefore represents a suitable alternative to the standard determination methods.

Table 8. Time to result overview for the different parameters
Parameter Metod Time to result
Ethanol GC ∼5 minutes (preparation) + ∼5 minutes (GC)
Glycerol  GC ∼5 minutes (preparation) + ∼5 minutes (GC)
Water  Karl Fischer titration ∼ 5 minutes 
Hydrogen peroxide Permanganate titration ∼5 minutes
Kontakt

Metrohm Polska

ul. Centralna 27
05-816 Opacz-Kolonia

Kontakt