饮用水镍钴双金属玻碳电极分析平台
AN-V-224
2020-05
zh
Simultaneous determination in low ng/L range on the GC-RDE modified with a bismuth film
由于镍和钴的毒性和对人体健康的不利影响,须控制其在饮用水中的浓度。因此,欧盟立法规定饮用水中镍的限值为 20 µg/L。目前,世界卫生组织的《饮用水水质指南》中镍的暂定指导值为最高浓度 70 微克/升。使用 884 专业 伏安极谱仪监测镍和钴的浓度时,采用的是在玻璃碳电极(GC-RDE)上同时测定镍和钴浓度的方法,该电极由 Bi 薄膜修饰。
Due to the toxicity and the detrimental effects of nickel and cobalt on human health, their concentrations in drinking water must be controlled. Therefore, EU the legislation specifies 20 µg/L as the limit value for nickel in drinking water. The current provisional guideline value for Ni in the World Health Organization’s «Guidelines for Drinking-water Quality» is set to a maximum concentration of 70 µg/L. To monitor the concentrations of Ni and Co with the 884 Professional VA, a method for simultaneous determination on the glassy carbon electrode (GC-RDE) modified with a Bi film is used.
The simplicity of the bismuth film preparation step allows quick and easy regeneration of the sensing layer. The determination is based on adsorptive stripping voltammetry of both elements using dimethylglyoxime (DMG) as a complexing agent. This method results in excellent performance in terms of sensitivity, reaching a limit of detection of 0.05 μg/L for Ni and 0.03 μg/L for Co. This non-toxic method is best suited for both manual and automated systems, allowing the determination in sample series with low to medium number of samples.
Drinking water, mineral water, sea water
Prior to the first determination, an ex-situ bismuth film is deposited from a bismuth solution. In the next step, the electrodes are cleaned with ultrapure water and the bismuth solution is removed. The water sample is placed into the measuring vessel. Ammonium buffer together with the complexing agent (DMG) are added, and the simultaneous determination of nickel and cobalt is carried out with the 884 Professional VA using the parameters specified in Table 1. The concentration is determined by two additions of a nickel and cobalt standard addition solution.
| Parameter | Setting |
|---|---|
| Mode | DP – Differential Pulse |
| Deposition potential | -0.8 V |
| Deposition time | 30 s |
| Start potential | -0.85 V |
| End potential | -1.25 V |
| Peak potential Ni | -0.97 V |
| Peak potential Co | -1.12 V |
- Working electrode: Glassy carbon (GC-RDE)
- Reference electrode: Ag/AgCl/KCl (3 mol/L)
- Auxiliary electrode: Glassy carbon rod
With a 30 s deposition time, this method is suitable for the determination of both nickel and cobalt in water samples in concentrations from β(Ni) = 0.2–8 µg/L and β(Co) = 0.1–10 µg/L. Using a 90 s deposition time, the limit of detection can be lowered to approximately 0.05 µg/L for nickel and 0.03 µg/L for cobalt.
| Sample | Ni (μg/L) | Co (μg/L) |
|---|---|---|
| Tap water | 0.34 | < LOD |