重新導向訊息

The guideline value for total chromium in the World Health Organization’s (WHO) «Guidelines for Drinking-water Quality» is 50 µg/L. Chromium(VI) is more toxic than its trivalent form (Cr(III)) and is also less abundant. Therefore a robust and sensitive method is required to monitor its concentration in drinking water.

The mercury film modified scTRACE Gold can be used to monitor chromium(VI), offering easy handling and a high grade of stability. The sensor itself contains all necessary electrodes: the gold micro-wire working electrode, Ag/AgCl reference, and the carbon auxiliary electrode integrated on a ceramic substrate which does not need intensive maintenance like polishing. Adsorptive stripping voltammetry (AdSV) using diethylenetriaminepentaacetic acid (DTPA) as a complexing agent and catalytic signal enhancement allows chromium(VI) detection at very low levels, with a limit of detection of 2 µg/L. The ability to replate the mercury film allows a quick and easy regeneration of the sensor. This method is best suited for field applications.

Drinking water, mineral water, sea water

946 Portable VA Analyzer
Figure 1. 946 Portable VA Analyzer

Prior to the first determination, the ex-situ mercury film is deposited on the scTRACE Gold electrode. In the next step, the electrodes are cleaned with ultrapure water and the measuring vessel is emptied. The water sample, the supporting electrolyte with the complexing agent (DTPA) are pipetted into the measuring vessel. The determination of chromium(VI) is carried out with a 946 Portable VA Analyzer using the parameters specified in Table 1. The concentration is determined by two additions of a chromium(VI) standard addition solution.

The scTRACE Gold is electrochemically activated prior to the first determination.

Table 1. Parameters
Parameter Setting
Mode DP – Differential Pulse
Start potential -1.15 V
End potential -1.65 V
Peak potential Cr(Vl) -1.4 V
  • scTRACE Gold

The method is suitable for the determination of chromium(VI) concentrations up to 40 µg/L. The limit of detection without deposition time is approximately 5 µg/L.

Figure 2. Determination of chromium(VI) in tap water spiked with 30 µg/L
Table 2. Result
Sample Cr(VI) (μg/L)
Tap water spiked with 30 µg/L Cr(VI) 32.1
聯絡我們

Metrohm AG

Ionenstrasse
9100 Herisau

聯絡我們

Internal reference: AW VA CH4-0596-042020