Byli jste přesměrováni na vaši místní verzi požadované stránky

In an acidic solution (containing NH4F * HF, Al(NO3)3 / KNO3) sodium forms NaK2AlF6 which precipitates in an exothermic solution, enabling analysis by thermometric titration. Several foods were analyzed, namely bouillon, gravy, tomato ketchup, corn chips, pretzel sticks as well as crackers. The reproducibility of the results was good. After weighing in and adding solutions, samples were crushed with a polytron to ensure homogeneity in the measuring solution. Relative standard deviations were between 0.08% and 3.75%. In addition to this application bulletin, you can find more information on thermometric sodium determination in foods in our application video:

Kontakt

Metrohm Česká republika s.r.o.

Na Harfě 935/5c
190 00 Praha

Kontakt